The relation between wedge factors in air and water

Raquel D. Bar-Deroma, Bengt E. Biarngard

Research output: Contribution to journalArticlepeer-review

Abstract

The wedge factor (WF) for an x-ray beam is defined as the ratio of doses with and without the wedge at a point on the central axis, in water. It is a function of depth d, field size s, and collimator setting c. It is demonstrated that WF can be calculated as WF(rf,5, c)=WFairAH,H;(c)/ A0(d)H0(c), where WFair is a constant factor, namely, the ratio of the incident fluences measured “in air” with and without the wedge for reference conditions, e.g., at isocenter for a 10X10 cm2 collimator setting. Aw(d) and A0(d) are attenuation factors, and Hw(c) and H0(c) head-scatter factors, with and without the wedge and normalized to reference conditions. This expression is based on the approximation that the scatter factor is the same with and without the wedge. The field size, which affects the scatter factor, is therefore not an important variable for the wedge factor using this approach. The above relation has been evaluated at 6 and 25 MV for various depths (d25 cm) and field sizes (s25 cm) for built-in wedges (wedge angle approximately 60°). The agreement was better than ±1.5%. This formulation of the wedge factor is easily implemented in calculations of dose and monitor settings.

Original languageEnglish
Pages (from-to)1043-1047
Number of pages5
JournalMedical Physics
Volume21
Issue number7
DOIs
StatePublished - Jul 1994
Externally publishedYes

Keywords

  • attenuation factor
  • head-scatter factor
  • scatter factor
  • wedge factor
  • wedge filters

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'The relation between wedge factors in air and water'. Together they form a unique fingerprint.

Cite this