Abstract
Rapid, highly sensitive, and high-throughput detection of biomarkers at low concentrations is invaluable for early diagnosis of various diseases. In many highly sensitive immunoassays, magnetic beads are used to capture fluorescently labeled target molecules. The target molecules are then quantified by detecting the fluorescent signal from individual beads, which is time consuming and requires a complicated and expensive detection system. Here, we demonstrate a high-throughput optical modulation biosensing (ht-OMB) system, which uses a small permanent magnet to aggregate the beads into a small detection volume and eliminates background noise by steering a laser beam in and out of the cluster of beads. Shortening the aggregation, acquisition, and well-to-well scanning transition times enables reading a 96-well plate within 10 min. Using the ht-OMB system to detect human Interleukin-8, we demonstrated a limit of detection of 0.14 ng/L and a 4-log dynamic range. Testing 94 RNA extracts from 36 confirmed RT-qPCR SARS-CoV-2-positive patients (Ct≤40) and 58 confirmed RT-qPCR SARS-CoV-2-negative individuals resulted in 100% sensitivity and 100% specificity.
Original language | English |
---|---|
Article number | 123624 |
Journal | Talanta |
Volume | 248 |
DOIs | |
State | Published - 1 Oct 2022 |
Externally published | Yes |
Bibliographical note
Publisher Copyright:© 2022 Elsevier B.V.
Keywords
- COVID-19
- Fluorescence-based immunoassays
- Magnetic beads
- Magnetic biosensing
- Optical biosensors
- SARS-CoV-2
ASJC Scopus subject areas
- Analytical Chemistry